Carbon Dioxide Sensor - Hazardous Area Certified
Standard Series of InfraRed (IR) Gas Sensors
Download Technical Data Sheets (pdf) - Low Ranges up to 5% volume -TDS0003, High Ranges up to 100% volume - TDS0020
Features
- Measurement ranges - 0-500ppm up to 0-5% Volume & 0-100% Volume CO2
- Operating voltage 3 to 5 volts dc, nominal current = 60mA
- Provides 'raw' signal output for processing external to the sensor
- OEM board available containing the necessary processing electronics and LCD display
- Fast response
- Internal temperature signal output
- Temperature compensated detector elements
- Wide operating temperature range of -20 to +50 deg. C
- Industry standard size - 20mm diameter x 16.6mm high (excluding pins)
- Optional replaceable, self-adhesive micro porous 1 micron filter
Description
Carbon Dioxide sensors within the Standard series of infrared gas sensors provide a 'raw' signal output which needs to be be fed into external processing electronics and software. This may either be provided by the equipment manufacturer or by use of a purpose-designed OEM board available from Dynament Ltd.
The sensors are contained in an Ex d stainless steel housing and are hazardous area certified to IECEx, ATEX and UL standards. The IECEx and ATEX certification also includes Group I (mining) certification. Further information.
These sensors can be used to measure Carbon Dioxide gas within a wide range of commercial and industrial applications. The standard measurement ranges that are available are summarised in the table below. If you have a requirement for a particular measurement range not shown below then please contact us and we will be please to assist you with your application.
Sensor Order Codes
The sensor order codes commence with the code MSH-CO2/TC (for sensors up to 5% volume) or MSH-HCO2/TC (for sensors up to 100% volume) followed by other options such as a filter and/or isolated case - please refer to the data sheets above for full details.
OEM Board for use with Standard Series Sensors
Dynament can provide this board for original equipment manufacturers who do not wish to develop their own electronics and software. It provides a ready built and tested micro-controller based circuit board for use with its sensors.
The micro-controller contains all the necessary software and signal processing algorithms to enable the user to minimise the time-to-market of their product/s. Custom software development facilities are also available.
Order Code OEM-1